Sale

Edwards nEXT400T HC Turbo Pump Package

Starting at: $18,439.00 Starting at: $13,829.25

The Edwards nEXT400T HC Turbo pump package comes with the turbo pump, TAG controller, 200W Tag power supply, fan cooling kit and connection cables. Note, gauge not included.

The nEXT400 turbomolecular pump builds upon the success of the highly successful EXT line to deliver a pump with significantly improved performance in a smaller package. The rotor has been totally re-designed to optimize pump performance and achieve both higher speeds and higher compression while maintaining high levels of reliability. In addition to enhanced performance nEXT400 benefits from the same service and operational functionalities from the larger nEXT range, including field serviceability and in-built intelligence, for a truly class leading product.

•  Class leading performance in a reduced footprint
•  Enhanced customer choice with standard and compression focused variants
•  Increased operating envelope maximized by Edwards’ patented direct impeller temperature measurement technology
•  Additional flexibility from customized interstage and backing port position selection
•  4 year service with full end-user field serviceability
•  CMSD enabled low vibration levels
•  Intelligent pumps with included micro USB port and Edwards nST software allowing pump monitoring, control and configuration
•  Full suite of tailored accessories

Please call for lead time.
Original price was: $18,439.00.Current price is: $13,829.25.
SKU: B83300300
Original price was: $20,228.00.Current price is: $15,171.00.
SKU: B83300400
View cart

Description

Edwards nEXT400T HC Turbo Pump Package

The Edwards nEXT400T HC Turbo pump package comes with the turbo pump, TAG controller, 200W Tag power supply, fan cooling kit and connection cables. Note, gauge not included.

The nEXT400 turbomolecular pump builds upon the success of the highly successful EXT line to deliver a pump with significantly improved performance in a smaller package. The rotor has been totally re-designed to optimize pump performance and achieve both higher speeds and higher compression while maintaining high levels of reliability. In addition to enhanced performance nEXT400 benefits from the same service and operational functionalities from the larger nEXT range, including field serviceability and in-built intelligence, for a truly class leading product.

•  Class leading performance in a reduced footprint
•  Enhanced customer choice with standard and compression focused variants
•  Increased operating envelope maximized by Edwards’ patented direct impeller temperature measurement technology
•  Additional flexibility from customized interstage and backing port position selection
•  4 year service with full end-user field serviceability
•  CMSD enabled low vibration levels
•  Intelligent pumps with included micro USB port and Edwards nST software allowing pump monitoring, control and configuration
•  Full suite of tailored accessories

Please call for lead time.

Additional information

Weight N/A
Edwards nEXT 400 Turbo Pkg

,

Brand

Edwards

Edwards is a leading developer and manufacturer of sophisticated vacuum products, abatement solutions and related value-added services.
Inlet flange DN160ISO-K, DN160CF
Outlet flange DN25NW
Pumping speed
  N2 400 ls-1
  He 390 ls-1
  H2 325 ls-1
Compression ratio D / T
  N2 >1 × 1011
  He 1 x 108  / >1 x 108
  H2 5 x 105 / 1 x 106
Ultimate pressure with RV backing pump ISO/CF <1 x 10-8 / <1 x 10-10 mbar
Recommended backing pump* RV12/nXDS10i
Vent port 1/8 inch BSP female
Purge port 1/8 inch BSP female
Max continuous inlet flow (Nitrogen)†
Water cooling (water at 15°C, ambient temp at 40°C) 105 sccm
Forced air cooled, 35 °C ambient 90 sccm
Pump rotational speed
Nominal rotational speed 60000 rpm
Standby rotational speed Variable from 33000 to 60000 rpm (42000 rpm default)
Programmable power limit settings Variable from 50-200W (160W default)
Start time to 90% speed D / T  160 W 175s / 200s
Analogue outputs Pump rotational speed;
Power consumption; Pump temp; Controller temp
Cooling method# Forced air / water
Ambient air temp for forced air cooling 5 – 35°C
Min cooling water flow rate (water 15°C) 15 l h-1
Water temp range 10 – 20°C
Max inlet flange temp 70°C
Operating attitude Vertical and upright, through to
Horizontal +/- 2°
Noise level at 1 metre <45 dB(A)
Max magnetic field pump can tolerate 5 mT
Recommended controller TIC200 turbo and instrument controller
Quiescent electrical power Typically 20 W