Edwards Active Pirani Gauges
Starting at: $547.00 Starting at: $533.33
Edwards Active Pirani Gauges
Edwards APG100 series Active Pirani vacuum gauges are available in 2 models. The APG100-XM is the standard model and measures to 10-3 mbar, the APG100-XLC is a corrosion resistant version with measurement to 10-4 mbar.
Both gauges feature compact size for easy installation, a linear output and a replaceable sensor tube. These gauges are compatible with all Edwards TIC instrument controllers and other Active gauge controllers and displays. They are also CSA, C/US approved as well as fully RoHS compliant due to their lead-free construction.
• Cable connections and gauge adjustment conveniently located, thereby minimizing the space envelope required for access
• Sensor tube can be baked to 150C
• Adjustable set-point for simple process control and interlocking
• Remote calibration possible
• NW16, NW25 and DN16CF flange options for easy connection to vacuum systems
• Linear output – 1 Volt per decade for easy interface with vacuum control systems
Choose from a KF16 or KF25 Version
- Description
- Additional information
- Brand
- Technical Specs
- Downloads
- Videos
Description
Edwards Active Pirani Gauges
Edwards APG100 series Active Pirani vacuum gauges are available in 2 models. The APG100-XM is the standard model and measures to 10-3 mbar, the APG100-XLC is a corrosion resistant version with measurement to 10-4 mbar.
Both gauges feature compact size for easy installation, a linear output and a replaceable sensor tube. These gauges are compatible with all Edwards TIC instrument controllers and other Active gauge controllers and displays. They are also CSA, C/US approved as well as fully RoHS compliant due to their lead-free construction.
• Cable connections and gauge adjustment conveniently located, thereby minimizing the space envelope required for access
• Sensor tube can be baked to 150C
• Adjustable set-point for simple process control and interlocking
• Remote calibration possible
• NW16, NW25 and DN16CF flange options for easy connection to vacuum systems
• Linear output – 1 Volt per decade for easy interface with vacuum control systems
Choose from a KF16 or KF25 Version
Additional information
| Weight | N/A |
|---|---|
| Edwards Pirani APG | APG 100-XLC KF16, APG 100-XLC KF25, APG 100-XM KF16, APG 100-XM KF25, APG200-XLC-NW16, APG200-XLC-NW25, APG200-XM-NW16, APG200-XM-NW25 |
Brand
| Mass | 85 g |
| Internal volume | 5 cm3 |
| Enclosure rating | IP40 |
| Measurement Range – APG100-XM | Atmosphere to 10-3 mbar |
| (APG100-XLC) | Atmosphere to 10-4 mbar |
| Accuracy (APG100-XM) | Typically +/- 15% at <100 mbar |
| Accuracy (APG100-XLC) | Typically +/- 15% at <10 mbar |
| Maximum over-pressure | 10 bar absolute |
| Operating temp range | 5o to 60o C |
| Storage temp range | -30o to 70 oC |
| Bake-out with no electronics | 150C |
| Humidity | 80% RH up to 31C decreasing linearly to 50% RH at 40C and above |
| Maximum altitude | 3000 m |
| Filament temperature | 100 oC above ambient |
| Electrical supply voltage | 15 to 30 V d.c. nominal / 13.5 V d.c. minimum / 32 V d.c. maximum |
| Power consumption | 1 W |
| Output signal | 0 to 10 V d.c. nominal |
| Set-point – open collector transistor | |
| Â Rating | 30 V d.c. 100 mA |
| Â Range of set-point | 1.8 to 9.2 V dc |
| Â Fixed hysteresis | 500 mV (1/2 decade) |
| Â Level setting resolution | 6 mV |
Edwards APG 100 Instruction Manual
Edwards APG 100 Replacement Tube and Electronic Housing Calibrating Instructions